Search Our Site
 
 
Equipment - Readers

ELX800LDI - MICROPLATE READER

As A Stand-Alone Microplate Reader, The ELx800 Is Designed For Applications Within The Clinical, Biotechnology Research And Pharmaceutical Areas. The On-Board Data Reduction Surpasses Many PC Software Packages With Its Extensive Curve-Fitting, Cutoff Calculation, Data Transformation And Validation Capabilities. The ELX800 Interfaces With All Of Bio-Tek's Software Packages Including KC4 And KC Junior.

 
  • Extensive on-board data analysis:
    • Curve-fitting options including linear, quadratic and 4-P.
    • Control and assay validation with complete results reporting.
    • Transformation formulas.
    • Cutoff and call criteria.
  • Memory:
    • Stores 55 assay definitions in memory (up to 75 custom assays).
    • Saves 8 microplate test results.
    • Stores up to 25 standard curves for reuse.
  • Multiple Plate Formats:
    • 384-well plate capability for High Throughput Screening applications.
    • Routine ELISA's and cellular assays in 24, 48, and 96 well analysis.
    • Specialized applications in 60, 72, and 96 well Terasaki formats.
  • On-Board Diagnostic Testing:
    • System Test checks the entire optics path.
    • Universal Test Plate software tests for alignment, accuracy, linearity and repeatability.
  • Interfaces To Many Robotic Workstations (optional).

  • Specifications:
    Format:
    • 1 Reader.
    Plate Type:
    • 24, 48, and 96-well strips.
    Speed:
    • 30 seconds for 96 wells.
    Filters:
    • 5-cavity wheel supplied: 405, 450, 490, 630 nm.
    Absorbance Range:
    • 0.000 to 3.000 Abs.
    Wavelength Range:
    • 400 to 750 nm.
    Resolution:
    • 0.001 Abs.
    Accuracy:
    • +/- 1% +/- 0.010 Abs from 0.000 to 2.000 Abs.
    Linearity:
    • +/- 1% from 0.000 to 2.000 Abs.
     
    • +/- 3% from 2.000 to 3.000 Abs.
    Repeatability:
    • +/- 0.5% +/- 0.005 Abs from 0.000 to 2.000 Abs.
    Dimensions:
    • 16.5" D x 15" W x 7" H (41.9 x 38.1 x 17.8 cm).
    Weight:
    • 18.5 lb. (8 kg.).
    External Outputs:
    • RS 232 Serial Interface.
     
    • Parallel Printer interface.
    Display:
    • 2x 24 LCD Display.
    Regulatory:
    • CE Mark, ETL Mark for UL3101-1, CAN/CSA C22.2 No. 1010.1, Conforms to IEC 1010-1.
    Read Method:
    • End point.
    Product ID #:
    • Q9800.