ENVIROLOGIX
Inc. - GMO Monitoring Kits
CRY1C
- PLATE KIT (ELISA Kit)
| |
Intended Use
|
The EnviroLogix Cry1C Plate Kit Is Designed
For The Quantitative Laboratory Detection Of Cry1C Residues
In Plant Leaf Tissue Samples.
| |
General
Use |
|
The
EnviroLogix Cry1C Plate Kit is a “sandwich” Enzyme-Linked
ImmunoSorbent Assay (ELISA). In the test, plant leaf
sample extracts are added to test wells coated with
antibodies raised against Cry1C toxin. Any residues
present in the sample extract bind to the antibodies,
and are then detected by addition of enzyme (horseradish
peroxidase) labeled Cry1C antibody. After a simple
wash step, the results of the assay are visualized with
a color development step; color development is proportional
to Cry1C concentration in the sample extract.
Lighter
color = Lower concentration
Darker
color = Higher concentration

| Materials
Provided: |
Materials Required
But Not Provided: |
|
12 x |
Strips
Of 8 Antibody-Coated Wells each, in plate
frame |
|
1 x |
Vial
Of Cry1C Negative Control. |
|
1 x |
Vial
Of 1 ppb Cry1C Calibrator. |
|
1 x |
Vial
Of 5 ppb Cry1C Calibrator. |
|
1 x |
Vial
Of 10 ppb Cry1C Calibrator. |
|
1 x |
Bottle Of Cry1C-Enzyme Conjugate. |
|
1 x |
Bottle
Of 5x Extraction/ Dilution Buffer. |
|
1 x |
Packet
Of Wash Buffer Salts. |
|
1 x |
Bottle
Of Substrate. |
|
1 x |
Bottle Of Stop Solution |
|
|
Disposable
Tissue Extractors, EnviroLogix Cat. # ACC
002 (6 samples are provided in the kit). |
|
|
Test
tubes for dilution of sample extracts. |
|
Disposable
Tip, Adjustable Air-Displacement Pipettes
which will measure 20, 100, 500 and 1000 micro
liters (µL). |
|
|
Marking
Pen (indelible). |
|
Tape or Parafilm. |
|
|
Timer
(15 minutes, 1 hour, and 30 minutes). |
|
Distilled
Or Deionized Water for preparing Wash Buffer
and diluting 5x Extraction/Dilution Buffer. |
|
|
Glass
Bottles or flasks with 175 mL capacity for
storage of 1x Extraction/Dilution Buffer and
1 liter capacity for Wash Buffer. |
|
Microtiter
Plate Reader or strip reader. |
|
|
Wash Bottle, or microtiter plate or strip
washer. |
|
Multi-Channel
Pipette that will measure 100 µL (optional). |
|
|
Racked
Dilution Tubes for loading samples into the
plate with a multichannel pipette (optional). |
|
Orbital
Plate Shaker (optional).
|
|
| Kit
Specifications: |
| Format:
|
|
| Time
Requirement: |
|
| Detection
Limit: |
- To
be determined (the lowest calibrator in the
kit is 1 ppb).
|
| Cross-Reaction:
(Limit
of Detection) |
- Cry1C
(1 ppb).
-
Cry1Ab (5000 ppb).
-
Cry1Ac (>10 000 ppb).
-
Cry2A (>10 000 ppb).
-
Cry1F (>10 000 ppb).
-
Cry9C (>10 000 ppb).
|
| Measurement:
|
|
| Product
ID #: |
|
|