Intended Use
|
The
EnviroLogix QualiPlate™ Kit for Cry1F is designed
for the qualitative laboratory detection of Cry1F endotoxin
in corn grain, seed or leaf samples.Two assay protocols
are presented: The High Sensitivity Protocol will detect
the Cry1F protein found in 0.25% Herculex I corn (1
positive kernel in 400 conventional kernels) and requires
2.5 hours to run. The Rapid Protocol (1.5 hours total)
is intended for use in screening individual seeds or
leaf samples for the presence of Cry1F.
| |
General
Use |
The
EnviroLogix QualiPlateTM Cry1F Kit is a “sandwich”
Enzyme-Linked ImmunoSorbent Assay (ELISA). In the test,
corn product sample extracts are added to test wells
coated with antibodies raised against Cry1F protein.
Any Cry1F present in the sample extract binds to the
antibodies, and is then detected by addition of enzyme
(horseradish peroxidase) labeled Cry1F antibody. After
a simple wash step, the results of the assay are visualized
with a color development step; color development is
proportional to Cry1F concentration in the sample extract.
Lighter
color = Lower concentration
Darker
color = Higher concentration

| Materials
Provided: |
Materials Required
But Not Provided: |
|
|
1
x |
|
Antibody-Coated
Solid Plate |
|
|
1
x |
|
Vial
Of Cry1F Positive Control. |
|
|
1
x |
|
Bottle
Of Cry1F Enzyme Conjugate. |
|
|
1
x |
|
Packet
Of Wash Buffer Salts. |
|
|
1
x |
|
Bottle
Of Substrate. |
|
|
1
x |
|
Bottle Of Stop Solution |
|
|
Distilled Or Deionized Water for preparing
wash buffer. |
| |
Glass
Bottles or flasks plus graduated cylinder
with one liter capacity for preparation and
storage of Wash and Extraction Buffer |
|
Tween-20
(Sigma P-1379, or Equivalent) for preparation
of Extraction Buffer. |
|
|
Waring
laboratory blender (model 31BL91 or equivalent)
for Preparation of Extraction Buffer |
|
Test
Or Centrifuge Tubes for extraction of grain
and dilution of sample extracts (optional). |
| |
Centrifuge capable of 5000 x g, (optional). |
|
Disposable Tip, Adjustable Air-Displacement
Pipettes which will measure 100 microliters
(µL). |
| |
Marking Pen (indelible). |
|
Tape
or Parafilm. |
| |
Timer. |
|
Microtiter
Plate Reader or strip reader. |
| |
Wash Bottle, or microtiter plate or strip
washer. |
|
Multi-Channel Pipette that will measure 100
µL (optional). |
| |
Racked Dilution Tubes for loading samples
into the plate with a multichannel pipette
(optional). |
|
Orbital Plate Shaker (optional). |
| |
Snap-cap
Tubes and pestles for extraction of leaf samples
(EnviroLogix Cat# ACC 002, 100/package), optional. |
|
|
| Kit
Specifications: |
| Format:
|
|
| Time
Requirement: |
- Rapid
Protocol: 1 h 30 min.
-
High Sensitivity Protocol: 2 h 30 min.
|
| Detection
Limit: |
-
High Sensitivity Protocol: 0.070 ppb.
|
| Measurement:
|
|
| Product
ID #: |
|
|
|