ENVIROLOGIX
Inc. - GMO Monitoring Kits
CRY9C
- PLATE KIT (ELISA Kit)
| |
Intended Use
|
The EnviroLogix Quantiplate Cry9C Kit
Is Designed For The Quantitative Or Qualitative Laboratory
Detection Of Cry9C Endotoxin In Corn Grain Or Seed.
The Most Prevalent Cry9C Varieties In Corn Are Starlink™
And Starlink Licensees. For Ease Of Reference Only,
Corn Modified With Cry9C Is Referred To Throughout This
Product Insert As Starlink. Two Assay Protocols Are
Presented: High Sensitivity And Rapid Protocols. Please
See The Application Guides In The Product Insert For
Instructions On Use Of This Kit With Corn Leaves, Single
Corn Seeds, Corn Flour, Cornmeal, And Corn Grits.
| |
General
Use |
|
The
EnviroLogix QuantiPlateTM Cry9C Kit is a “sandwich”
Enzyme-Linked ImmunoSorbent Assay (ELISA). In the test,
corn product sample extracts are added to test wells
coated with antibodies raised against Cry9C toxin.
Any residues present in the sample extract bind to the
antibodies, and are then detected by addition of enzyme
(horseradish peroxidase) labeled Cry9C antibody. After
a simple wash step, the results of the assay are visualized
with a color development step; color development is
proportional to Cry9C concentration in the sample extract.
Lighter
color = Lower concentration
Darker
color = Higher concentration

| Materials
Provided: |
Materials Required
But Not Provided: |
|
|
12
x |
|
Strips
Of 8 Antibody-Coated Wells each, in plate
frame |
|
|
1
x |
|
Vial
Of Cry9C Negative Control. |
|
|
1
x |
|
Vial
Of 0.01% StarLink (0.2 ppb Cry9C) Calibrator. |
|
|
1
x |
|
Vial
Of 0.04% StarLink (0.8 ppb Cry9C) Calibrator. |
|
|
1
x |
|
Vial
Of 0.125% StarLink (2.5 ppb Cry9C) Calibrator. |
|
|
1
x |
|
Vial Of 0.25% StarLink (5 ppb Cry9C) Calibrator. |
|
|
1
x |
|
Bottle
Of Cry9C-Enzyme Conjugate. |
|
|
1
x |
|
Bottle
20x Grain Extraction Concentrate |
|
|
1
x |
|
Packet
Of Wash Buffer Salts. |
|
|
1
x |
|
Bottle
Of Substrate. |
|
|
1
x |
|
Bottle Of Stop Solution |
|
|
Distilled Or Deionized Water for preparing
wash buffer and diluting 20x grain extraction
concentrate). |
|
|
Glass
Bottles or flasks with 1.2 litre capacity
for preparation of 1x Grain Extraction Solution
and 1 liter capacity for preparation of Wash
Buffer, plus graduated cylinders for measuring
the components of these solutions. |
|
Test
Or Centrifuge Tubes for extraction of grain
and dilution of sample extracts (optional). |
|
|
Centrifuge capable of 5000 x g, or low protein-binding
hydrophilic syringe filters, 0.45 um (such
as Pall Gelman Sciences Product #
4184) (optional). |
|
Disposable Tip, Adjustable Air-Displacement
Pipettes which will measure 100 and 1000 micro
liters (µL) and 5 ml. |
|
|
Marking Pen (indelible). |
|
Tape
or Parafilm. |
|
|
Timer. |
|
Microtiter
Plate Reader or strip reader. |
|
|
Wash Bottle, or microtiter plate or strip
washer. |
|
Multi-Channel Pipette that will measure 100
µL (optional). |
|
|
Racked Dilution Tubes for loading samples
into the plate with a multichannel pipette
(optional). |
|
Orbital Plate Shaker (optional). |
|
| Kit
Specifications: |
| Format:
|
|
| Time
Requirement: |
- Rapid
Protocol: 2 h 15 min.
-
High Sensitivity Protocol: 3 h 45 min.
|
| Detection
Limit: |
-
Rapid Protocol: 0.250 ppb.
-
High Sensitivity Protocol: 0.070 ppb.
|
| Quantification
Limit: |
-
Rapid Protocol: 6 ppb.
-
High Sensitivity Protocol: 1.5 ppb.
|
Cross-Reaction:
(Limit of Detection) |
-
Cry9C (0.070 ppb).
-
Cry1Ab (>50,000 ppb).
-
Cry1Ac (>50,000 ppb).
-
Cry2A (>50 000 ppb).
-
Cry1F (66 ppb).
-
Cry1C (>50,000 ppb).
|
| Measurement:
|
|
| Product
ID #: |
|
|
|