EnviroLogix
Inc.- Environmental & Food Monitoring Kits: Pesticides
ACETANILIDES
(ELISA Kit)
| |
Intended Use
|
The
EnviroLogix Acetanilide Plate Kit Is Designed For The
Quantitative Laboratory Detection Of Acetanilide Pesticide
Residues In Ground And Surface Water Samples, With An
Assay Range From 0.04 To 1.0 Parts Per Billion (Ppb).
| |
General |
|
The
EnviroLogix Acetanilide Plate Kit is a competitive Enzyme-Linked
ImmunoSorbent Assay (ELISA). In the test, Acetanilide
pesticide residues in the sample compete with enzyme
(horseradish peroxidase) labeled alachlor for a limited
number of antibody binding sites on the inside surface
of the test wells. After a simple wash step, the outcome
of the competition is visualized with a color development
step. As with all competitive immunoassays, sample
concentration is inversely proportional to color development.
Darker
color = Lower concentration
Lighter
color = Higher concentration The
EnviroLogix Acetanilide Plate Kit can be utilized in
applications such as:
- Source
water protection.
- Treated
effluent monitoring.
- Drinking
water monitoring.
- Run-off
assessment & monitoring.
-
Point source testing.
| Materials
Provided: |
Materials Required
But Not Provided: |
|
8 x |
Strips
of 12 Antibody-Coated Wells each, in plate
frame. |
|
1 x |
Vial
of Negative Control. |
|
1 x |
Vial
of 0.04 ppb Alachlor Calibrator. |
|
1 x |
Vial
of 0.2 ppb Alachlor Calibrator. |
|
1 x |
Vial
of 1.0 ppb Alachlor Calibrator. |
|
1 x |
Bottle
of Alachlor-Enzyme Conjugate. |
|
1 x |
Bottle
of Substrate. |
|
1 x |
Bottle
of Stop Solution. |
|
|
Disposable
Tip, Adjustable Air Displacement Pipette which
will measure 150 microliters (µL), 100
µL and 50 µL. |
|
|
Marking
Pen (indelible). |
|
Tape
or Parafilm. |
|
|
Timer
(1 hour and 30 minutes). |
|
Cool
Tap or Distilled Water. |
|
|
Microtiter Plate Reader or strip reader. |
|
Microtiter Plate Washer (optional). |
|
|
Twelve-Channel Pipette that will measure 150
µL, 100 µL and 50 µL (optional). |
|
Racked
Dilution Tubes (optional). |
|
|
Orbital
Plate Shaker (optional).
|
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Limit
Of Detection: |
|
| Limit
of Quantification: |
|
Cross
Reaction: (Limit
Of Detection) |
-
Alachlor (0.03 ppb).
- Alachlor
Mercapturate (0.05 ppb).
-
Alachlor ESA (0.16 ppb).
-
Acetochlor (0.06 ppb).
-
Acetochlor Mercapturate (5.3 ppb).
-
Acetochlor ESA (18 ppb).
-
Metalaxyl (7.1 ppb).
-
Metolachlor (0.06 ppb).
-
Metolachlor Mercapturate (3.6 ppb).
-
Metolachlor ESA (83 ppb).
-
Propachlor (923 ppb).
|
| Standards: |
|
| Measurement: |
-
Microtiter Plate Reader or strip reader.
|
| Product
ID #: |
|
|