Search Our Site
 
 

ENVIROLOGIX Inc. - Human Exposure Monitoring Kits

ALACHLOR MERCAPTURATE (ELISA Kit)

Intended Use  

The EnviroLogix Alachlor Mercapturate In Urine Plate Kit Is Designed For The Quantitative Laboratory Detection Of Alachlor Mercapturate Residues In Human Urine Samples, With A Quantitative Assay Range From 20 To 650 Parts Per Billion (Ppb).

General

 

The EnviroLogix Alachlor Mercapturate in Urine Plate Kit is a competitive Enzyme-Linked ImmunoSorbent Assay (ELISA).  In the test, Alachlor Mercapturate residues in the sample compete with enzyme (horseradish peroxidase)-labeled Alachlor for a limited number of antibody binding sites on the inside surface of the test wells.  After a simple wash step, the outcome of the competition is visualized with a color development step.  As with all competitive immunoassays, sample concentration is inversely proportional to color development.

Darker color = Lower concentration

Lighter color = Higher concentration

The EnviroLogix Alachlor Mercapturate in Urine Plate Kit can be used in the following testing applications:

  • Human Exposure Monitoring
  • Public Health Screening Programs
  • Health Risk Assessment

Materials Provided: Materials Required But Not Provided:
  • 8 x   Strips of 12 Antibody-Coated Wells each, in plate frame
  • 1 x    Vial of Negative Control
  • 1 x     Vial of 20 ppb Alachlor Mercapturate Calibrator (Actual Concentration = 0.4 ppb*)
  • 1 x   Vial of 125 ppb Alachlor Mercapturate Calibrator (Actual Concentration = 2.5 ppb*).
  • 1 x    Vial of 650 ppb Alachlor Mercapturate Calibrator (Actual Concentration = 13 ppb*).
  • 1 x   Bottle of Alachlor-Enzyme Conjugate.
  • 1 x   Bottle of 10 x Sample Diluent
  • 1 x   Bottle of Substrate.
  • 1 x   Bottle of Stop Solution.
        *  The difference between the stated and actual concentration is to correct for the 1:50 dilution required for sample preparation.
  • Disposable Tip, Adjustable Air Displacement Pipette, which will measure 980 microliters (µL), 150 µL, 100 µL, 50 µL and 20 µL.
  • Glass Test Tubes.
  • Marking Pen (indelible).
  • Tape or Parafilm.
  • Timer (1 hour and 30 minutes).
  • Distilled Water.
  • Cool Tap Or Distilled Water for rinsing wells.
  • Cool Tap Or Distilled Water for rinsing wells.
  • Microtiter Plate Reader or strip reader.
  • Microtiter Plate Washer (optional).
  • Twelve-Channel Pipette that will measure 50 µL, 100 µL and 150 µL (optional).
  • Racked Dilution Tubes for loading samples into the plate with a 12-channel pipette (optional).
  • Orbital Plate Shaker (optional).

    Kit Specifications:
    Plate Format: 
    • 96 Well.
    Time Requirement: 
    • 2 h 15 min.
    Detection Limit:
    • 10 ppb.
    Quantification Limit:     
    • 25 ppb.
    Cross Reaction:
    (Limit Of Detection)
    • Alachlor Mercapturate (10 ppb).
    • Atrazine Mercapturate (>10000 ppb).
    • Acetochlor Mercapturate (>1000 ppb).
    • Metolachlor Mercapturate (135 ppb).
    • Alachlor ESA (115 ppb).
    • Acetochlor ESA (>10000 ppb).
    • Metolachlor ESA (>10000 ppb).
    Measurement: 
    • Microtiter Plate Reader or strip reader.
    Product ID #:
    • BP001.