| ENVIROLOGIX
Inc. - Human Exposure Monitoring Kits
ALACHLOR
MERCAPTURATE (ELISA Kit)
| |
Intended Use
|
The
EnviroLogix Alachlor Mercapturate In Urine Plate Kit
Is Designed For The Quantitative Laboratory Detection
Of Alachlor Mercapturate Residues In Human Urine Samples,
With A Quantitative Assay Range From 20 To 650 Parts
Per Billion (Ppb).
| |
General |
|
The
EnviroLogix Alachlor Mercapturate in Urine Plate Kit
is a competitive Enzyme-Linked ImmunoSorbent Assay (ELISA).
In the test, Alachlor Mercapturate residues in the sample
compete with enzyme (horseradish peroxidase)-labeled
Alachlor for a limited number of antibody binding sites
on the inside surface of the test wells. After a simple
wash step, the outcome of the competition is visualized
with a color development step. As with all competitive
immunoassays, sample concentration is inversely proportional
to color development.
Darker
color = Lower concentration
Lighter
color = Higher concentration
The
EnviroLogix Alachlor Mercapturate in Urine Plate Kit
can be used in the following testing applications:
-
Human Exposure Monitoring
- Public
Health Screening Programs
- Health
Risk Assessment

| Materials
Provided: |
Materials Required
But Not Provided: |
|
|
8
x |
|
Strips
of 12 Antibody-Coated Wells each, in plate
frame |
|
|
1
x |
|
Vial
of Negative Control |
|
|
1
x |
|
Vial of 20 ppb Alachlor Mercapturate Calibrator
(Actual Concentration = 0.4 ppb*) |
|
|
1
x |
|
Vial
of 125 ppb Alachlor Mercapturate Calibrator
(Actual Concentration = 2.5 ppb*). |
|
|
1
x |
|
Vial
of 650 ppb Alachlor Mercapturate Calibrator
(Actual Concentration = 13 ppb*). |
|
|
1
x |
|
Bottle
of Alachlor-Enzyme Conjugate. |
|
|
1
x |
|
Bottle
of 10 x Sample Diluent |
|
|
1
x |
|
Bottle
of Substrate. |
|
|
1
x |
|
Bottle
of Stop Solution. |
| |
|
* |
The
difference between the stated and actual concentration
is to correct for the 1:50 dilution required
for sample preparation. |
|
| |
Disposable
Tip, Adjustable Air Displacement Pipette,
which will measure 980 microliters (µL),
150 µL, 100 µL, 50 µL and
20 µL. |
| |
Glass
Test Tubes. |
| |
Marking
Pen (indelible). |
| |
Tape or Parafilm. |
| |
Timer (1 hour and 30 minutes). |
|
Distilled
Water. |
| |
Cool Tap Or Distilled Water for rinsing wells.
|
| |
Cool Tap Or Distilled Water for rinsing wells. |
| |
Microtiter
Plate Reader or strip reader. |
| |
Microtiter Plate Washer (optional). |
|
Twelve-Channel
Pipette that will measure 50 µL, 100
µL and 150 µL (optional). |
| |
Racked
Dilution Tubes for loading samples into the
plate with a 12-channel pipette (optional). |
|
Orbital
Plate Shaker (optional).
|
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
Cross
Reaction: (Limit
Of Detection) |
-
Alachlor Mercapturate (10 ppb).
-
Atrazine Mercapturate (>10000 ppb).
-
Acetochlor Mercapturate (>1000 ppb).
-
Metolachlor Mercapturate (135 ppb).
-
Alachlor ESA (115 ppb).
-
Acetochlor ESA (>10000 ppb).
-
Metolachlor ESA (>10000 ppb).
|
| Measurement: |
-
Microtiter Plate Reader or strip reader.
|
| Product