ENVIROLOGIX
Inc. - Environmental & Food Monitoring Kits: Pesticides
ALDICARB (ELISA Kit)
| |
Intended Use
|
The
EnviroLogix Aldicarb Plate Kit Is Designed For The Quantitative
Laboratory Detection Of Aldicarb Pesticide Residues
In Ground And Surface Water Samples, With An Assay Range
From 3 To 100 Parts Per Billion (Ppb).
| |
General
Use |
|
The
EnviroLogix Aldicarb Plate Kit is a competitive Enzyme-Linked
ImmunoSorbent Assay (ELISA). In the test, Aldicarb
pesticide residues in the sample compete with enzyme
(horseradish peroxidase) labeled Aldicarb for a limited
number of antibody binding sites on the inside surface
of the test wells. After a simple wash step, the outcome
of the competition is visualized with a color development
step. As with all competitive immunoassays, sample
concentration is inversely proportional to color development.
Darker
color = Lower concentration
Lighter color = Higher concentration
The
EnviroLogix Aldicarb Plate Kit can be used in the following
applications:
- Source
Water Protection.
- Treated
Effluent Monitoring.
-
Drinking Water Monitoring.
- Run-Off
Assessment And Monitoring.
- Point
Source Testing.
- Soil Testing.

| Materials
Provided: |
Materials Required
But Not Provided: |
|
1 x |
Antibody-Coated
Microwell Plate; 8 removable strips of 12
microwells each, in recloseable foil bag with
desiccant. |
|
1 x |
Negative
Control. |
|
1 x |
3
ppb Aldicarb Calibrator. |
|
1 x |
20
ppb Aldicarb Calibrator. |
|
1 x |
100
ppb Aldicarb Calibrator. |
|
1 x |
Aldicarb
- Enzyme Conjugate. |
|
1 x |
Substrate. |
|
1 x |
Stop
Solution. |
|
|
Distilled
Water (or equivalent). |
|
|
Glass
Vials or Tubes. |
|
Positive-Displacement
Pipettor with disposable tips and plungers
(capable of delivering 50 microliters (µL)). |
|
|
Pipette
capable of delivering 5 milliliters (mL). |
|
|
Disposable
Tip, Adjustable Air Displacement Pipette which
will measure 50 µL, 100 µL and
150 µL. |
|
|
Marking
Pen (indelible). |
|
|
Tape
or Parafilm. |
|
|
Timer
(1 hour and 30 minutes). |
|
|
Microtiter
Plate Reader or strip reader. |
|
|
Microtiter
Plate Washer (optional). |
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
Cross
Reaction: (Detection
limit) |
- Aldicarb
(2.6 ppb).
- Aldicarb-Sulfone
(580 ppb).
-
Aldicarb Sulfoxide (3950 ppb).
-
Atrazine (270 ppb).
|
| Measurement: |
-
Microtiter Plate Reader or Strip Reader.
|
| Product
ID #: |
|
|