ENVIROLOGIX
Inc. - Environmental & Food Monitoring Kits: Pesticides
CHLORPYRIFOS (ELISA Kit)
| |
Intended Use
|
The
EnviroLogix Chlorpyrifos Plate Kit Is Designed For The
Quantitative Laboratory Detection Of Chlorpyrifos Pesticide
Residues In Ground And Surface Water Samples, With An
Assay Range From 0.3 To 6 Parts Per Billion (Ppb).
| |
General |
|
The
EnviroLogix Chlorpyrifos Plate Kit is a competitive
Enzyme-Linked ImmunoSorbent Assay (ELISA). In the test,
Chlorpyrifos pesticide residues in the sample compete
with enzyme (horseradish peroxidase) labeled Chlorpyrifos
for a limited number of antibody binding sites on the
inside surface of the test wells. After a simple wash
step, the outcome of the competition is visualized with
a colour development step. As with all competitive
immunoassays, sample concentration is inversely proportional
to color development.
Darker
color = Lower concentration
Lighter color = Higher concentration
The EnviroLogix
Chlorpyrifos Plate Kit can be used in the following
applications:
- Source
Water Protection.
- Treated
Effluent Monitoring.
- Drinking
Water Monitoring.
- Run-Off
Assessment & Monitoring.
- Point
Source Testing.
| Materials
Provided: |
Materials Required
But Not Provided: |
|
|
8
x |
|
Strips
of 12 Antibody-Coated Wells each, in plate
frame.
|
|
|
1
x |
|
Vial
of 1 microgram (µg) per milliliter (mL)
Chlorpyrifos Calibrator Stock Solution (in
methanol). |
|
|
1
x |
|
Bottle
of Chlorpyrifos-Enzyme Conjugate. |
|
|
1
x |
|
Bottle of Substrate. |
|
|
1
x |
|
Bottle
of Stop Solution. |
|
|
Distilled Water (or equivalent). |
|
|
Glass
Vials or tubes. |
|
Positive-Displacement
Pipetter with disposable tips and plungers
(capable of delivering 60 microliters (µL)). |
|
|
Pipette(s) capable of delivering 10, 9.5,
7.5, 2.5 and 0.5 mL. |
|
Disposable
Tip, Adjustable Air Displacement pipette which
will measure 50 µL, 100 µL and
150 µL. |
|
|
Marking
Pen (indelible). |
|
Tape
or Parafilm. |
|
|
Timer
(1 hour and 30 minutes). |
|
Cool
Tap or Distilled Water. |
|
|
Microtiter
Plate Reader or strip reader. |
|
Microtiter
Plate Washer (optional). |
|
|
Twelve-Channel
Pipette that will measure 50 µL, 100
µL and 150 µL (optional). |
|
Racked
Glass Dilution Tubes (optional). |
|
|
Orbital
Plate Shaker (optional).
|
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
| Cross
Reaction: |
- Chlorpyrifos
(0.3 ppb).
-
Chlorpyrifos-Methyl (4.1 ppb).
-
Diazinon (6.5 ppb).
-
Quinalphos (17 ppb).
-
Azinphos-Ethyl (21 ppb).
-
Azinphos-Methyl (43 ppb).
-
Pirimiphos-Ethyl (33 ppb).
-
Pirimiphos-Methyl (87 ppb).
-
Parathion (32 ppb).
-
Parathion-Methyl (165 ppb).
-
Fenitrothion (95 ppb).
-
Malathion (233 ppb).
|
| Measurement: |
- Microtiter
Plate Reader or Strip Reader.
|
| Product
ID #: |
|
|