ENVIROLOGIX Inc. -
Environmental & Food Monitoring Kits: Pesticides
CYANAZINE
(ELISA Kit)
| |
Intended Use
|
The EnviroLogix Cyanazine Plate Kit Is
Designed For The Quantitative Laboratory Detection Of
Cyanazine Pesticide Residues In Ground And Surface Water
Samples, With An Assay Range From 0.4 To 10 Parts Per
Billion (Ppb).
| |
General |
|
The EnviroLogix
Cyanazine Plate Kit is a competitive Enzyme-Linked ImmunoSorbent
Assay (ELISA). In the test, Cyanazine pesticide residues
in the sample compete with enzyme (horseradish peroxidase)
labeled Cyanazine for a limited number of antibody binding
sites on the inside surface of the test wells. After
a simple wash step, the outcome of the competition is
visualized with a color development step. As with all
competitive immunoassays, sample concentration is inversely
proportional to color development.

Darker
color = Lower concentration
Lighter
color = Higher concentration
The
EnviroLogix Cyanazine Plate Kit can be used in the following
applications:
- Source Water Protection.
- Treated Effluent Monitoring.
- Drinking Water Monitoring.
- Run-Off Assessment & Monitoring.
- Point Source Testing.
- Soil Testing.

| Materials
Provided: |
Materials Required
But Not Provided: |
|
8 x |
Strips
of 12 Antibody-Coated Wells each, in plate
frame. |
|
1 x |
Vial
of Negative Control. |
|
1 x |
Vial of 0.4 ppb Cyanazine Calibrator. |
|
1 x |
Vial
of 2.0 ppb Cyanazine Calibrator.
|
|
1 x |
Vial
of 10.0 ppb Cyanazine Calibrator. |
|
1 x |
Bottle
of Cyanazine-Enzyme Conjugate.. |
|
1 x |
Bottle
of Substrate. |
|
1 x |
Bottle
of Stop Solution. |
|
|
Disposable Tip, Adjustable Air Displacement
Pipette which will measure 100 microliters
(µL). |
|
|
Marking
Pen (indelible). |
|
Tape or Parafilm. |
|
|
Timer
(1 hour and 30 minutes). |
|
Cool
Tap or Distilled Water for rinsing wells. |
|
|
Microtiter
Plate Reader or strip reader. |
|
Microtiter
Plate Washer (optional). |
|
|
Twelve-Channel
Pipette that will measure 100 µL (optional). |
|
Racked
Dilution Tubes for loading samples into the
plate with a 12-channel pipette (optional). |
|
|
Racked Dilution Tubes for loading samples
into the plate with a 12-channel pipette (optional).
|
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
| Cross
Reaction: |
-
Cyanazine (0.14 ppb).
-
Ametryn (36 ppb).
-
De-Ethylated Atrazine (2.2 ppb).
-
Atrazine (3.3 ppb).
-
Prometon (86 ppb).
-
Prometryn (57 ppb).
-
Propazine (2.9 ppb).
-
Simazine (60 ppb).
-
Simetryn (188 ppb).
-
Terbuthylazine (0.04 ppb).
|
| Measurement: |
-
Microtiter Plate Reader or strip reader.
|
| Product