| ENVIROLOGIX
Inc. - Environmental & Food Monitoring Kits: Pesticides
DDT (ELISA Kit)
| |
Intended
Use |
The EnviroLogix DDT Plate Kit Is A Quantitative
Laboratory Assay For The Detection Of Total DDT Pesticide
Residues (Sum Of P,P’-DDT, -DDE, And -DDD) In Soil,
With An Assay Range From 0.025 To 0.875 Parts Per Million
(Ppm).
| |
General |
|
The
EnviroLogix DDT Plate Kit is a competitive Enzyme-Linked
ImmunoSorbent Assay (ELISA). In the test, DDT pesticide
residues in the sample compete with enzyme (horseradish
peroxidase) labeled DDT for a limited number of antibody
binding sites on the inside surface of the test wells.
After a simple wash step, the outcome of the competition
is visualized with a color development step. As with
all competitive immunoassays, sample concentration is
inversely proportional to color development.
Darker
color = Lower concentration
Lighter
color = Higher concentration
The EnviroLogix
DDT Plate Kit can be used in the following applications:
- Source Water Protection.
- Treated Effluent Monitoring.
- Drinking Water Monitoring.
- Run-Off Assessment & Monitoring.
- Crop And Food Testing.
- Soil Testing.

| Materials
Provided: |
Materials Required
But Not Provided: |
|
12 x |
Strips
of 8 Antibody-Coated Wells each, in plate
frame 1 vial of Negative Control (methanol). |
|
1 x |
Vial
of “0.025 ppm” DDT Calibrator
(actual concentration 5 ppb in methanol*). |
|
1 x |
Vial
of “0.1 ppm” DDT Calibrator (actual
concentration 20 ppb in methanol*).
|
|
1 x |
Vial
of “0.875 ppm” DDT Calibrator
(actual concentration 175 ppb in methanol*). |
|
1 x |
Bottle
of DDT Assay Diluent. |
|
1 x |
Bottle
of DDT-enzyme Conjugate Concentrate (100x). |
|
1 x |
Bottle
of Substrate.
|
|
1 x |
Bottle
of Stop Solution. |
| * |
The
reason for the difference between the stated
and actual calibrator concentrations is to
account for the 1:5 dilution incurred during
sample extraction. |
|
|
Balance
For Weighing Samples |
|
|
EnviroLogix Soil Extraction Kit, or suitable
sealable vessels for sample extraction. |
|
Methanol
for sample extraction |
|
|
Positive
displacement pipetter with disposable tips
and plungers for dispensing of calibrators
and sample extracts (capable of delivering
20 microliters (µL)). |
|
Disposable Tip, Adjustable Air Displacement
Pipette(s) which will measure 120, 60 and
100 µL. |
|
|
Marking Pen (indelible). |
|
Tape or Parafilm. |
|
|
Timer (1 hour and 30 minutes). |
|
Cool
Tap or Distilled Water for rinsing wells. |
|
|
Microtiter
Plate Reader or strip reader. |
|
Microtiter
Plate Washer (optional). |
|
|
Multi-Channel
Pipette that will measure 120, 60 and 100
µL (optional). |
|
Orbital
Plate Shaker (optional). |
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
| Measurement: |
-
Microtiter Plate Reader or Strip Reader.
|
| Product
ID #: |
|
|