| ENVIROLOGIX
Inc. - Environmental & Food Monitoring Kits: Pesticides
FLUOMETURON (ELISA Kit)
| |
Intended Use
|
The
EnviroLogix Fluometuron Plate Kit Is Designed For The
Quantitative Laboratory Detection Of Fluometuron Pesticide
Residues In Surface Water Samples, With An Assay Range
From 0.05 To 3.5 Parts Per Billion (Ppb). Ground Water
Samples May Be Analyzed, But Will Require The Addition
Of A Chelating Agent. (For Information Concerning Ground
Water Samples Please Contact Xygen Diagnostics Inc.)
| |
General |
|
The
EnviroLogix Fluometuron Plate Kit is a competitive Enzyme-Linked
ImmunoSorbent Assay (ELISA). In the test, Fluometuron
pesticide residues in the sample compete with enzyme
(horseradish eroxidase) labeled Fluometuron for a limited
number of antibody binding sites on the inside surface
of the test wells. After a simple wash step, the outcome
of the competition is visualized with a color development
step. As with all competitive immunoassays, sample
concentration is inversely proportional to color development.
Darker
color = Lower concentration
Lighter
color = Higher concentration
The
EnviroLogix Fluometuron Plate Kit can be used in the
following applications:
- Run-Off Assessment & Monitoring:
- Surface Water (Ponds, Retention Pools, Streams,
Lakes).
- Ground Water.

| Materials
Provided: |
Materials Required
But Not Provided: |
|
8 x |
Strips
of 12 Antibody-Coated Wells each, in plate
frame. |
|
1 x |
Vial
Negative Control. |
|
1 x |
Vial
of 0.05 ppb Fluometuron Calibrator (in deionized
water). |
|
1 x |
Vial
of 0.5 ppb Fluometuron Calibrator (in deionized
water). |
|
1 x |
Vial
of 3.5 ppb Fluometuron Calibrator (in deionized
water). |
|
1 x |
Bottle
of Fluometuron-Enzyme Conjugate. |
|
1 x |
Bottle
of Substrate. |
|
1 x |
Bottle
of Stop Solution. |
|
|
Disposable
Tip Adjustable Air Displacement Pipette which
will measure 100 µL. |
|
|
Marking
Pen (indelible). |
|
Tape
or Parafilm. |
|
|
Timer
(1 hour and 30 minutes). |
|
Cool
Tap or Distilled Water for rinsing wells. |
|
|
Microtiter
Plate Reader or strip reader. |
|
Microtiter
Plate Washer (optional). |
|
|
Twelve-Channel
Pipette that will measure 100 µL (optional). |
|
Racked Dilution Tubes for loading samples
into the plate with a 12-channel pipette (optional). |
|
|
Orbital
Plate Shaker (optional).
|
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
Cross
Reaction: (Limit
Of Detection) |
- Fluometuron
(0.03 ppb).
- Chlortoluron
(0.28 ppb).
-
Monuron (15 ppb).
-
Diuron (0.13 ppb).
-
Linuron (1.2 ppb).
-
Isoproturon (1355 ppb).
|
| Measurement: |
- Microtiter
Plate Reader or Strip Reader.
|
| Product
ID #: |
|
|