ENVIROLOGIX
Inc. - Environmental & Food Monitoring Kits - Natural Toxins
MICROCYSTIN
- QuantiPlateTM Plate Kit (ELISA Kit)
| |
Intended Use
|
The EnviroLogix
QuantiPlateTM Microcystin Kit Is Designed For The
Quantitative Laboratory Detection Of Microcystin Toxin In
Surface Water Samples, With An Assay Quantitation Range From
0.16 To 2.5 Parts Per Billion (Ppb). (See The Appendix At
The End Of The Package Insert Describing An Alternate Assay
Protocol With A Quantitation Range From 0.05 To 0.83 Ppb.)
| |
General |
|
The EnviroLogix
QuantiPlateTM Microcystin Kit is a competitive
Enzyme-Linked ImmunoSorbent Assay (ELISA). In the test, Microcystin
toxin in the sample competes with enzyme (horseradish peroxidase)
labeled Microcystin for a limited number of antibody binding
sites on the inside surface of the test wells. After a simple
wash step, the outcome of the competition is visualized with
a color development step. As with all competitive immunoassays,
sample concentration is inversely proportional to color development.
Darker
color = Lower concentration
Lighter
color = Higher concentration
The EnviroLogix
Microcystin Plate Kit can be used in the following testing
applications:
- Early Warning
Of Bloom Formation And Water Quality Degradation.
- Ground
And Surface Water Screening And Monitoring.
- Ensure
Guidance Levels Are Not Exceeded In Drinking Water, Surface
Water, And Algae Based Food Supplements.
- Quality
Control Testing Of Raw Material And Finished Algae Based
Products.
| Materials
Provided: |
Materials Required
But Not Provided: |
|
12 x |
Strips
of 8 Antibody-Coated Wells each, in plate frame. |
| 1
x |
Vial
of Negative Control. |
| 1
x |
Vial
of 0.16 ppb Microcystin LR Calibrator.
|
| 1
x |
Vial
of 0.6 ppb Microcystin LR Calibrator |
| 1
x |
Vial
of 2.5 ppb Microcystin LR Calibrator. |
| 1
x |
Bottle
of Assay Diluent. |
| 1
x |
Bottle
of Microcystin-Enzyme Conjugate |
| 1
x |
Packet
of Wash Solution Salts. |
| 1
x |
Bottle
of Substrate |
| 1
x |
Bottle
of Stop Solution |
|
|
Disposable
Tip Adjustable Air-Displacement pipetter which will
measure 20 µL, 100 µL and 125 µL. |
| |
Marking
Pen (indelible). |
|
Tape
or Parafilm. |
| |
Timer
(30 minutes). |
|
Distilled
Water for preparing Wash Solution. |
| |
Glassware
for storing Wash Solution |
|
Wash
Bottle for washing strips with Wash Solution. |
| |
Microtiter
Plate Reader or strip reader. |
|
Microtiter
Plate Washer (optional). |
| |
Twelve-Channel
Pipette that will measure 20 µL, 100 µL
and 125 µL (optional). |
| |
Racked
Dilution Tubes for loading samples into the plate
with a 12-channel pipette (optional). |
| |
Orbital
Plate Shaker (optional).
|
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
Cross
Reaction:
(Limit
Of Detection) |
- Microcystin
LR (0.15 ppb).
-
Microcystin LA (0.24 ppb).
-
Microcystin RR (0.27 ppb).
-
Microcystin YR (0.44 ppb).
-
Nodularin (0.21 ppb).
|
| Measurement: |
- Microtiter
Plate Reader or Strip Reader.
|
| Product
ID #: |
|
|