| EnviroLogix
Inc. - Environmental & Food Monitoring Kits: Pesticides
PARAQUAT
(ELISA Kit)
| |
Intended Use
|
The EnviroLogix Paraquat Plate Kit Is
Designed For The Quantitative Laboratory Detection Of
Paraquat Pesticide Residues In Ground And Surface Water
Samples, With An Assay Quantitation Range From 0.04
To 1.2 Parts Per Billion (Ppb).
| |
General |
|
The EnviroLogix Paraquat
Plate Kit is a competitive Enzyme-Linked ImmunoSorbent
Assay (ELISA). In the test, Paraquat pesticide residues
in the sample compete with enzyme (horseradish peroxidase)
labeled Paraquat for a limited number of antibody binding
sites on the inside surface of the test wells. After
a simple wash step, the outcome of the competition is
visualized with a color development step. As with all
competitive immunoassays, sample concentration is inversely
proportional to color development.
Darker color = Lower concentration
Lighter color = Higher concentration
The EnviroLogix Paraquat Plate Kit can
be used in the following applications:
- Source Water Protection.
- Drinking Water Monitoring.
- Treated Effluent Monitoring.
- Run-Off Assessment And Monitoring.
- Point Source Testing.
- Soil Testing.

| Materials
Provided: |
Materials Required
But Not Provided: |
|
8 x |
Strips
of 12 Antibody-Coated Wells each, in plate
frame. |
|
1 x |
Vial
Negative Control. |
|
1 x |
Vial
of 0.04 ppb Paraquat Calibrator (in deionized
water). |
|
1 x |
Vial
of 0.2 ppb Paraquat Calibrator (in deionized
water). |
|
1 x |
Vial
of 1.2 ppb Paraquat Calibrator (in deionized
water). |
|
1 x |
Bottle
of Paraquat-Enzyme Conjugate.
|
|
1 x |
Bottle
of Substrate. |
|
1 x |
Bottle
of Stop Solution. |
|
|
Disposable
Tip Adjustable Air Displacement Pipette which
will measure 100 µL. |
|
|
Marking Pen (indelible). |
|
Tape
or Parafilm. |
|
|
Timer
(1 hour and 30 minutes). |
|
Cool
Tap or Distilled Water for rinsing wells. |
|
|
Microtiter
Plate Reader or strip reader. |
|
Microtiter Plate Washer (optional). |
|
|
Twelve-Channel
Pipette that will measure 100 µL (optional). |
|
Racked
Dilution Tubes for loading samples into the
plate with a 12-channel pipette (optional). |
|
|
Orbital Plate Shaker (optional). |
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
| Cross
Reaction: |
-
Paraquat (0.01 ppb).
-
Diquat (300 ppb).
|
| Measurement: |
- Microtiter
Plate Reader or Strip Reader.
|
| Product