| ENVIROLOGIX
Inc. - Environmental & Food Monitoring Kits: Pesticides
PARATHION (ELISA Kit)
| |
Intended Use
|
The EnviroLogix Parathion Plate Kit
Is Designed For The Quantitative Laboratory Detection
Of Parathion Pesticide Residues In Ground And Surface
Water Samples, With An Assay Range From 0.04 To 4 Parts
Per Billion (Ppb).
| |
General |
The
EnviroLogix Parathion Plate Kit is a competitive Enzyme-Linked
ImmunoSorbent Assay (ELISA). In the test, Parathion
pesticide residues in the sample compete with enzyme
(horseradish peroxidase) labeled Parathion for a limited
number of antibody binding sites on the inside surface
of the test wells. After a simple wash step, the outcome
of the competition is visualized with a color development
step. As with all competitive immunoassays, sample
concentration is inversely proportional to color development.
Darker
color = Lower concentration
Lighter
color = Higher concentration
The
EnviroLogix Parathion Plate Kit can be used in the following
applications:
- Source Water Protection.
- Drinking Water Monitoring.
- Treated Effluent Monitoring.
- Run-Off Assessment And Monitoring.
- Crop And Food Testing.
- Soil Testing.

| Materials
Provided: |
Materials Required
But Not Provided: |
|
8 x |
Strips
of 12 Antibody-Coated Wells each, in plate
frame. |
|
1 x |
Vial
of 1000 ppb Parathion Calibrator Stock Solution
(in methanol). |
|
1 x |
Bottle
of Parathion-enzyme conjugate. |
|
1 x |
Bottle
of Substrate. |
|
1 x |
Bottle
of Stop Solution. |
|
|
Distilled
Water (or equivalent) for preparation of calibrators. |
|
|
Glass
Vials or tubes for preparation of calibrators
(do not prepare or store calibrators in any
plastic containers). |
|
Positive-Displacement
Pipetter with disposable tips and plungers
for preparation of calibrators (capable of
delivering 20 microliters (µL)). |
|
|
Pipette(s)
capable of delivering 5, 4.95, 3.5, 0.5 and
0.05 mL. |
|
Disposable
Tip, Adjustable Air Displacement Pipette which
will measure 100 µL. |
|
|
Marking
Pen (indelible). |
|
Tape
or Parafilm. |
|
|
Timer
(30 minutes). |
|
Cool
Tap Or Distilled Water for rinsing wells. |
|
|
Micro Titer Plate Reader or strip reader.
|
|
Microtiter
Plate Washer (optional). |
|
|
Twelve-Channel
Pipette that will measure 100 µL (optional). |
|
Racked
Dilution Tubes for loading samples into the
plate with a 12-channel pipette (optional). |
|
|
Orbital
Plate Shaker (optional).
|
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
| Cross
Reaction: |
-
Parathion (0.023 ppb).
-
Coumaphos (0.013 ppb).
-
Quinalphos (0.4 ppb).
|
| Measurement: |
- Microtiter
Plate Reader or Strip Reader.
|
| Product
ID #: |
|
|