ENVIROLOGIX
Inc. - Environmental & Food Monitoring Kits: Pesticides
SYNTHETIC PYRETHOIDS (ELISA Kit)
| |
Intended Use |
This Kit Is Designed To Quantitate Synthetic
Pyrethroid Residues In Methanol Extracts Of Various
Samples. The Range Of The Assay Is From 4 To 80 Parts
Per Billion (Ppb) Of Cyfluthrin In Methanol Extracts.
These Cyfluthrin Calibrators May Be Used To Quantitate
Cyfluthrin, And To Approximate Deltamethrin, Cypermethrin,
And Lambda-Cyhalothrin Residues, As The Assay Cross-Reactivity
For These Compounds Is Nearly Equal. To Assay Quantitatively
For Other Synthetic Pyrethroids, Contact Xygen Diagnostics
For Alternative Calibrator Materials.
| |
General |
|
The
EnviroLogix Synthetic Pyrethroid Plate Kit is a competitive
Enzyme-Linked ImmunoSorbent Assay (ELISA). In the test,
synthetic pyrethroid pesticide residues in the sample
compete with enzyme horseradish peroxidase)-labeled
cypermethrin for a limited number of antibody binding
sites on the inside surface of the test wells. After
a simple wash step, the outcome of the competition is
visualized with a color development step. As with all
competitive immunoassays, sample concentration is inversely
proportional to color development.
Darker
color = Lower concentration
Lighter
color = Higher concentration
The
EnviroLogix Synthetic Pyrethoid Plate Kit can be used
in the following applications:
Grain Storage Protection.
Drinking Water Monitoring.
Treated Effluent Monitoring.
Run-Off Assessment And Monitoring.
Source Water Protection.
Treatment Chemical Efficacy.

| Materials
Provided: |
Materials Required
But Not Provided: |
|
8 x |
Strips
of 12 Antibody-Coated Wells each, in plate
frame. |
|
1 x |
Vial
of Synthetic Pyrethroid Negative Control (neat
methanol). |
|
1 x |
Vial
of 4 ppb Cyfluthrin Calibrator, in methanol. |
|
1 x |
Vial
of 20 ppb Cyfluthrin Calibrator, in methanol. |
|
1 x |
Vial
of 80 ppb Cyfluthrin Calibrator, in methanol. |
|
1 x |
Vial
of Sodium Hydroxide Solution (0.2 N). |
|
1 x |
Bottle
of Cypermethrin-enzyme Conjugate.
|
|
1 x |
Bottle
of Substrate.
|
|
1 x |
Bottle
of Stop Solution. |
|
|
Methanol,
ACS Reagent Grade for sample extraction. |
|
|
Distilled
Or Deionized Water for calibrator and sample
extract dilution. |
|
Glass
Vials or test tubes for preparing dilutions. |
|
|
Positive-Displacement
Pipettor with disposable tips and plungers
capable of delivering 100 microliters (µL)
and 1 milliliter (mL). |
|
Repeating Pipettor, with syringe tip capable
of delivering 25 µL and 2 mL (also 3
syringe tips for delivering 100 µL if
you use the repeater to deliver conjugate,
substrate and stop in the assay). |
|
|
Vortex
Mixer. |
|
Disposable
Tip, Adjustable Air Displacement Pipette which
will measure 100 µL. |
|
|
Marking Pen. |
|
Tape
or Parafilm. |
|
|
Timer.
|
|
Cool
Tap Or Distilled Water for rinsing wells. |
|
|
Microtiter
Plate Reader or strip reader. |
|
Microtiter Plate Washer (optional). |
|
|
Twelve-Channel
Pipette that will deliver 100 µL (if
more than 3 strips are run per assay). |
|
Racked
Dilution Tubes for loading samples into the
plate with a 12-channel pipette (see previous
statement). |
|
|
Orbital
Plate Shaker (optional).
|
|
| Kit
Specifications: |
| Plate
Format: |
|
| Time
Requirement: |
|
| Detection
Limit: |
|
| Quantification
Limit: |
|
Cross
Reaction: (Limit
Of Detection) |
- Cyfluthrin
(2.5 ppb).
-
Deltamethrin (1.0 ppb).
-
Cypermethrin (1.7 ppb).
-
Lambda-Cyhalothrin (0.8 ppb).
|
| Measurement: |
- Microtiter
Plate Reader or Strip Reader.
|
| Product
ID #: |
|
|